1. The carbon dioxide incubator cannot be turned on before the water is injected, otherwise the heating element will be damaged.

2. After the incubator has been in operation for several months, the water in the water tank may be reduced due to volatilization. When the low water level indicator is on, it should be supplemented with water. First open the overflow pipe, use the funnel to connect the rubber tube to add water from the water injection hole to make the low water level indicator light off, then measure the supplementary water, and then block the overflow hole.

3. The carbon dioxide incubator can be used in a high-precision constant temperature incubator. At this time, the CO2 control system must be turned off.

4. Since the CO2 sensor is calibrated under saturated humidity, the humidifying disc must always be filled with sterilized water.

5. When the display temperature exceeds the set temperature by 1 °C, the over-temperature alarm indicator lights up and a sharp alarm sounds. At this time, the power should be turned off for 30 minutes; if the power supply (temperature control) switch is still over-temperature, the power should be turned off. And reported to the maintenance staff.

6. When the cylinder pressure is lower than 0.2MPa, the cylinder should be replaced.

7. Minimize the time to open the glass door.

8. If the carbon dioxide incubator is not used for a long time, the moisture in the working room must be removed before closing. Open the glass door and ventilate for 24 hours before closing.

9. When cleaning the CO2 incubator studio, do not hit the sensor and the stirring motor wind wheel and other components.

10, disassembly and assembly of the indoor support bracket, must use a special wrench, not excessive force.

11. The water in the tank must be removed before handling the incubator. When draining, put the rubber tube tightly on the water outlet hole so that the nozzle is lower than the instrument, gently suck a bit, and put down the water pipe, the water is siphone out.

12. Before carrying the carbon dioxide incubator, take out the shelf and humidifying tray in the working chamber to prevent the glass door from colliding.

13. When carrying the incubator, do not invert it. At the same time, do not lift the door to avoid deformation of the door.

Shangyu Xingxing Instrument Equipment Co., Ltd. is located in Daoxu Town, Shangyu District, Shaoxing City, Zhejiang Province. It is an enterprise integrating production, product research and development, processing and trade. The company's main products: drying box series, incubator series, laboratory muffle furnace series, distilled water series, as well as clean workbench, air shower, fume hood, biological safety cabinet and other complete purification equipment.